Description
This fully-loaded instrument is capable of working with all of the pipe and tube transduction systems produced by GUL. It is suitable for both the lower frequencies that are used for buried and coated pipes and the higher frequencies that are used with HD rings for support inspection and higher sensitivity to pitting. With one button press, the instrument can be placed into gPIMS Collector mode to allow a minimally trained operator to collect repeat data from a permanently installed gPIMS sensor.
The instrument contains 16 channels; therefore you will need a pair of channel reducers when joining EFC or HD inflatable rings together to collect data on very large diameter pipes.
Features:
Lightweight 4.5 kg / 9.9 lbs
High Speed Data Acquisition
16 Data Channels
On-Board Diagnostics Channel
7” LCD Touch Screen
Extended Frequency Range
Battery Operation
Operator Identification
Built-In GPS
Optional:
- Laptop computer with pre-installed WavePro™
- Wide range of LEMO cables
- EFC Processing Licence
- Absolute Calibration Software
- Ethernet Adaptor Cable (for LAN connection)
- In-car charger
- Warranty Extension Options
Specification
Designed for | All pipe and tube applications |
Dimensions | Approx. 22 x 30 x 13cm |
Weight (approx.) | 4.4 kg |
Transmitter Frequency Range | 7 to 150kHz |
Max output drive voltage | 400Vpp |
Max rep-rate | 20 Hz |
Convertor sampling frequency | 100 to 500kHz |
Receiving gain range | 10 to 90 dB |
Analog Band PAss Filter Range (Variable) | 7 to 150kHz |
Max sample range | 400m |
Max number of averages | 256 |
Battery type (Removable) | 6.6 Ah, 14.8V Li-Ion |
Typical battery life | At least 150 standard tests |
External power supply to charge instrument | 18-20 VDC (60W min) |
Number of transducer channels | 16 |
Screen | 7inch colour LCD touchscreen (800×480 pixels) |
Supported Operating Systems | Windows XP to Windows 7 |